Anisotropy of the device for each energy does not exceed values (in %)
presented in Table 1.1, when the device is rotated in the horizontal plane
(Appendix B, Figure B.1) and values (in %) presented in Table 1.2, when
the device is rotated in the vertical plane (Appendix B, Figure B.2).
25
0.059
0.662
1.25
0
0
0
0
15
±5
±5
±5
30
±10
±10
±10
45
±20
±15
±15
60
±40
±20
±20
-15
±5
±5
±5
-30
±10
±10
±10
-45
±20
±15
±15
-60
±40
±20
±20
Angle of the detection
relative to the direction
of the graduation
Energy of gamma radiation, MeV
Anisotropy
0.059
0.662
1.25
0
0
0
0
15
±5
±5
±5
30
±10
±10
±10
45
±20
±15
±15
60
±40
±20
±20
-15
±5
±5
±5
-30
±10
±10
±10
-45
±20
±15
±15
-60
±40
±20
±20
Angle of the detection
relative to the direction
of the graduation
Energy of gamma radiation MeV
Anisotropy
Table 1.1