Instruction Manual - Rev. 4
Model PB-5
Precision Pulse Generator
Berkeley Nucleonics Corporation 2955 Kerner Blvd., San Rafael, CA 94901
Phone: 415-453-9955, Fax: 415-453-9956, Email: info@berkeleynucleonics.com, Web: www.berkeleynucleonics.com
Page view 0
1 2 3 4 5 6 ... 18 19

Summary of Contents

Page 1 -

Instruction Manual - Rev. 4 Model PB-5 Precision Pulse Generator Berkeley Nucleonics Corporation 2955 K

Page 2 - WARRANTY

9 SECTION 2 RS-232 COMMANDS A PC can control all parameters of the model PB-5. The following commands will allow full operation of the pulser i

Page 3 - SECTION 1

11 SECTION 3 PB5 HOST INSTRUCTIONS 3.1 WHAT IS PB5 HOST? PB5 Host is a software program that provides BNC Model PB-5 Precision Pulse G

Page 4

12 7) If more than one Model PB-5 are connected, select the appropriate COM Port for each PB-5 Module (see Fig. 4) 8) Click the Connect check box

Page 5

13 Figure 4. Multiple PB-5 view 3.5 TOOL BAR NAVIGATION Menu Keyboard Shortcut Action File | Open Saved Setting Load previously saved setting

Page 6

14 3.6 COMMAND FORMAT [Channel]PB5_Command Parameter on Multi Channel Mode and Single Channel Mode. PB5_Command Parameter on Single Channel Mode

Page 7

15 Command Summary (-cont-) Command Keyboard Shortcut Parameter Range Description DELAy E & C 250n - 10m Set delay to parameter specified. Ct

Page 8

16 Command Summary (-cont-) Command Keyboard Shortcut Parameter Range Description TRIGger INternal, EXternal, GAted, ONe, FIre Set trigger to exter

Page 9

17 3.8 PB5 HOST SUPPORT Berkeley Nucleonics Corp. 2955 Kerner Blvd. San Rafael, CA 94901-5533 www.berkeleynucleonics.com Toll Free 800.234.78

Page 10 - SECTION 2

18 SECTION 4 SPECIFICATIONS Rep Rate: 2.0 Hz to 500 kHz Width: 100 ns to 1 ms, continuously variable Delay: 250 ns to 10 ms Rise Time

Page 11 - SECTION 3

19 Power Required: + 24 V 170 mA - 24 V 150 mA + 12 V 450 mA -12V 5 mA Mechanical: Triple-width AEC NIM module 4.05" wide x 8.70" high

Page 12

1 TABLE OF CONTENTS SECTION 1. OPERATIONS Page 1.1 Introduction

Page 13

2 SECTION 1 OPERATIONS 1.1 INTRODUCTION The PB-5 sets a new standard for nuclear pulse generators. It offers full digital controllabil

Page 14

3 FIGURE 1. Block Diagram of Model PB-5 Precision Pulse Generator The Trigger Out Generator provides a pulse synchronized with the beginning of the

Page 15

4 factors from 1 to 1000. Factory testing and calibration is provided via an external tester. A multi-pin connector separates the µC from the analog

Page 16

5 Hyper Terminal settings: Flow control: None Data bits: 8 Emulation: VT 100 Parity: None 1.4 MENU SELECTIONS Figure 2. shows the two level me

Page 17

6 FIGURE 2. Model PB-5 Menu 1.5 OPTIMUM PERFORMANCE To achieve optimum performance of the model PB-5 several factors should be conside

Page 18 - SECTION 4

7 used for low-amplitude tail pulses, the flat top portion will become longer as amplitude decreases. Since some amount of flat top is desir

Page 19

8 edges of each channel and calculating the width of each channel individually. This method is subject to error since it is difficult to