BUTSt:GATE:POLarity?
“NORM” or “INV” will be returned.
-- C --
CALibration?
Perform a calibration using the specified calibration value (CALibration:VALue
command). You need to enter a correct security code (see CALibration:
SECure:STATe) before you can calibrate the waveform generator.
CALibration:COUNt?
Query the number of times the waveform generator has been calibrated. Note
that your instrument was calibrated before it left the factory. Check the
calibration count when you receive your waveform generator to determine it
initial value.
CALibration:SECure:CODE <new code>
Set a new security code. You have to first use the old security code to unsecure
the waveform generator and then set a new security code. A security code may
contain up to 12 characters and the first character must be a letter (A-Z). The
rest of the characters can be any combination of numbers (0-9), letters and
underscore character (“_”).
CALibration:SECure:STATe {OFF|ON},<code>
Unsecure or secure the waveform generator for calibration.
CALibration:SECure:STATe?
Query the secured state of the waveform generator. Return “0” (OFF) or “1”
(ON).
CALibration:SETup <012394>
Configure the waveform generator’s internal state for each of the calibration
step to be performed.
CALibration:SETup?
Query the calibration setup number. Return a value from “0” to “94”.