allowed due to selection of pulse waveform.
-222 Data out of range; burst period; value clipped to…
The burst period is out of range and is adjusted.
-222 Data out of range; burst count; value clipped to…
The burst count is out of range and is adjusted.
-222 Data out of range; burst period limited by length of burst;
value clipped to lower limit
The burst period is too short to accommodate the burst count and
frequency and is adjusted to the minimum value allowed.
-222 Data out of range; burst count limited by length of burst;
value clipped to upper limit
The burst count is out of range. When the immediate trigger source is
selected, the burst count must be less than the product of the burst
period and the waveform frequency.
-222 Data out of range; amplitude; value clipped to…
The amplitude is out of range and is adjusted.
-222 Data out of range; offset; value clipped to…
The offset voltage is out of range and is adjusted.
-222 Data out of range; frequency in burst mode; value clipped
to…
The frequency is adjusted as indicated by the burst period.
-222 Data out of range; frequency in FM; value clipped to…
This message indicates that the carrier frequency is adjusted to the
lower boundary determined by the FM:DEV command. The carrier
frequency must be greater than or equal to the frequency deviation.
-222 Data out of range; marker confined to sweep span; value
clipped to…
The specified marker frequency is out of range and is adjusted to the