111
All non-volatile memory locations are occupied. There is no room to
store a new waveform.
781 Not enough memory to store new arb waveform; bad sectors
Due to storage hardware failure, new waveforms cannot be saved.
782 Cannot overwrite a built-in waveform
The five built-in waveform names (“EXP_RISE”, “EXP_FALL”,
“NEG_RAMP”, “SINC”, and “CARDIAC”) are reserved and cannot be
used with DATA:COPY command.
784 Name of source arb waveform for copy must be VOLATILE
When using the DATA:COPY command, the data source must be the
VOLATILE one.
785 Specified arb waveform does not exist
Before using DATA:COPY command to copy from the volatile memory,
be sure to download the waveform to it using the DATA VOLATILE or
DATA:DAC VOLATILE command.
786 Not able to delete a built-in arb waveform
Deleting the built-in waveform is not allowed.
787 Not able to delete the currently selected active arb
waveform
Deleting the waveform that is currently outputting is not allowed.
788 Cannot copy to VOLATILE arb waveform
You can only use DATA:COPY command to copy from VOLATILE, not to
VOLATILE.
800 Block length must be even
The binary data are represented as 16-bit integers. They are sent in
groups of two bytes (DATA:DAC VOLATILE command).
810 State has not been stored
The storage location specified in the *RCL command was never used in
Page view 110