command. You can set the dc level to any value between ±5 Vdc into 50
ohms or ±10 Vdc into an open circuit.
VOLTage? [MINimumMAXimum]
Query the output amplitude for the current function. The value is returned in the
unit chosen by the VOLT:UNIT command.
VOLTage:OFFSet {<offset>MINimumMAXimum}
Specify the dc offset voltage. The default is 0 volts. The minimum value is the
most negative dc offset for the chosen function and amplitude and the maximum
value is the largest dc offset for the chosen function and amplitude.
z The output amplitude and offset voltage are related to Vmax as shown
below:
|V
offset| + Vpp ÷ 2 Vmax
Where Vmax is the maximum peak voltage for the selected output termination
(5 volts for a 50Ω load or 10 volts for a high-impedance load). The function
generator will take the new specified output amplitude, but the offset voltage
may be reduced accordingly, and a “Settings conflict” error is generated.
z When the Output Termination Setting is changed, the offset limits are
automatically adjusted. For instance, when you change the output
termination from 50 ohms to “high impedance”, your offset voltage
doubles, thus it affects your output amplitude, and when Output
Termination has been changed from “high impedance” to 50 ohms, the
offset voltage decreases in half.
z For arbitrary waveforms, if the waveform data points do not span the full
range of output DAC (Digital-to-Analog Converter), the maximum offset
and amplitude will be limited.
z Setting a high level and low level is another way to specify the offset. For
instance, you will have an amplitude of 5 Vpp (with an associated offset
voltage of -500 mV) when you set your high level to +2 Vpp and low level
to -3 Vpp.
VOLTage:OFFSet? [MINimumMAXimum]
Query the dc offset voltage for the current function.
VOLTage:HIGH {<voltage>MINimumMAXimum}
Specify the high voltage level. The default high level for all functions is +50 mV.
Page view 165